FDA Adverse Event
Malfunction
Summary report: N
NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE
MDR report key: 2123326
·
Received June 8, 2011
Report
- Report Number
- 1811755-2011-02062
- Event Type
- Malfunction
- Date Received
- June 8, 2011
- Date of Event
- May 10, 2011
- Report Date
- May 10, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- FYD
- PMA / PMN Number
- K012991
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS EVALUATED BY A FIELD SERVICE REP AND THE EVENT WAS CONFIRMED. THE WATER SUPPLIED BY THE DOCKER WAS 185 TO 190 DEGREES FAHRENHEIT AND THE O-RING UNDER THE FLAPPER VALVE COULD NOT WITHSTAND THE HEAT AND ALLOWED THE WATER TO PASS DURING THE WASH CYCLE. ACCORDING TO THE NEPTUNE WASTE MANAGEMENT SYSTEM INSTRUCTIONS FOR USE THE WATER TEMPERATURE SHOULD BE 40 TO 110 DEGREES FAHRENHEIT. THE REPAIR TECHNICIAN NOTIFIED THE CUSTOMER TO ENSURE THAT THE COLD WATER WAS NOT TURNED OFF. THE DEVICE WAS REPAIRED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE FLAPPER VALVE ON THE NEPTUNE KEEPS OPENING DURING DOCKING AND SPITTING OUT FLUID. THERE WAS NO PT INVOLVEMENT AND NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE | FYD | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |