FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 2123319 · Received June 8, 2011

Report

Report Number
1811755-2011-02066
Event Type
Malfunction
Date Received
June 8, 2011
Date of Event
May 10, 2011
Report Date
May 10, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY A FIELD SERVICE REP AND THE EVENT WAS CONFIRMED. THE WATER SUPPLIED BY THE DOCKER WAS 185 TO 190 DEGREES FAHRENHEIT AND THE O-RING UNDER THE FLAPPER VALVE COULD NOT WITHSTAND THE HEAT AND ALLOWED THE WATER TO PASS DURING THE WASH CYCLE. ACCORDING TO THE NEPTUNE WASTE MANAGEMENT SYSTEM INSTRUCTIONS FOR USE, THE WATER TEMPERATURE SHOULD BE 40 TO 110 DEGREES FAHRENHEIT. THE REPAIR TECHNICIAN NOTIFIED THE CUSTOMER TO ENSURE THAT THE COLD WATER WAS NOT TURNED OFF. THE DEVICE WAS REPAIRED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE FLAPPER VALVE ON THE NEPTUNE KEEPS OPENING DURING DOCKING AND SPITTING OUT FLUID. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK