FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2123274 · Received June 14, 2011

Report

Report Number
2649622-2011-07952
Event Type
Malfunction
Date Received
June 14, 2011
Date of Event
March 7, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE DEFIBRILLATOR CONDUCTOR WAS DISTORTED, THE INNER TUBING WAS KINKED/BUCKLED, THE HELIX WAS DISTORTED/BENT, AND THE LEAD WAS DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE LEAD TOOK TOO MANY TURNS TO DEPLOY THE HELIX. THE PHYSICIAN IT THE LEAD FELT "FUNNY" AND THERE WAS HIGH IMPEDANCE WHEN IT WAS TESTED. THE LEAD WAS NOT USED. A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other