FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS MRI
MDR report key: 2123273
·
Received June 14, 2011
Report
- Report Number
- 2649622-2011-07951
- Event Type
- Malfunction
- Date Received
- June 14, 2011
- Date of Event
- March 3, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HELIX DID NOT EXTEND AFTER SEVERAL ATTEMPTS DURING THE LEAD PREPARATION. THE LEAD WAS ON STERILE TROLLEY. THE PHYSICIAN THOUGHT THE LEAD COULD BE IMPLANTED AND THUS USED A NEW LEAD. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Other |