FDA Adverse Event
Injury
Summary report: N
VIRTUOSO DR
MDR report key: 2123268
·
Received June 14, 2011
Report
- Report Number
- 6000144-2011-02259
- Event Type
- Injury
- Date Received
- June 14, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS FOUND THE CAUSE OF THE HIGH CURRENT DRAIN TO BE CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.
Description of Event or Problem · 1
THE DEVICE WAS REPLACED DUE TO ERI (ELECTIVE REPLACEMENT INDICATOR). THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D154AWG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Other | (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE TACHY LEAD |