FDA Adverse Event
Injury
Summary report: N
CONSULTA CRT-D
MDR report key: 2123259
·
Received June 14, 2011
Report
- Report Number
- 6000144-2011-02258
- Event Type
- Injury
- Date Received
- June 14, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S084
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE DEVICE WAS FULLY FUNCTIONAL WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE HAD SHORT BATTERY LIFE. IT WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONSULTA CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL, INC. | D224TRK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Hospitalization| R | 6947 IMPLANTABLE TACHY LEAD| 4196 IMPLANTABLE PACING LEAD| 4076 IMPLANTABLE PACING LEAD |