FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2123238 · Received June 14, 2011

Report

Report Number
2649622-2011-07940
Event Type
Injury
Date Received
June 14, 2011
Date of Event
February 27, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RECEIVED AND ANALYZED. THERE WERE NO ANOMALIES FOUND. ADDITIONAL FINDINGS OF THE PROXIMAL CONDUCTOR BLOOD/BODY FLUID (NOT OBSTRUCTED), AN OUTER INSULATION BREACHED CUT, AND BLOOD IN/ON HELIX MECHANISM ALONG WITH APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT FOLLOWING IMPLANT THERE WAS A PERFORATION OF THE INTERVENTRICULAR SEPTUM. THE LEAD WAS REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE OBSERVED AS A RESULT OF THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R