FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2123074 · Received June 14, 2011

Report

Report Number
2531779-2011-04122
Event Type
Malfunction
Date Received
June 14, 2011
Report Date
May 18, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP ON (B)(6) 2011 AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

FOLLOW-UP #2 - PRODUCT ANALYSIS: THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/22/2015 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THE UP KEYPAD BUTTONS WAS INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, THE BOLUS BUTTON WAS UNRESPONSIVE; CONTAMINATION WAS OBSERVED ON THE BOLUS BUTTON CONTACT; THE BOLUS BUTTON CONTACT WAS MISALIGNED. TWO BATTERY COMPARTMENT CRACKS WERE OBSERVED. THE RETURNED BATTERY CAP SECURED PROPERLY TO THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE PUMP AUDIO BOLUS BUTTON IS SLOW TO RESPOND AND REQUIRES EXCESSIVE FORCE TO ACTIVATE. THE PATIENT REPORTEDLY WORE THE PUMP IN A PROTECTIVE CASE AND DID NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 1200 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 57 YR