FDA Adverse Event Malfunction Summary report: N

ADVIA CENTAUR CP

MDR report key: 2122922 · Received June 14, 2011

Report

Report Number
2432235-2011-00085
Event Type
Malfunction
Date Received
June 14, 2011
Date of Event
May 13, 2011
Report Date
May 15, 2011
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Product Code
MMI
PMA / PMN Number
K902336
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS FIELD SERVICE ENGINEER WAS DISPATCHED TO THE CUSTOMER SITE. INSPECTION OF THE SYSTEM REVEALED A BUILD UP OF A WHITE PRECIPITATE AROUND THE PUMP 2 AND WD (WASH DISPLACEMENT) PORTS ON THE VALVE MANIFOLD. THE FSE REPLACED THE VALVE MANIFOLD AND RAN QC. THE INSTRUMENT HAS BEEN REPAIRED. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Additional Manufacturer Narrative · 1

A SIEMENS FIELD SERVICE ENGINEER WAS DISPATCHED TO THE CUSTOMER SITE. INSPECTION OF THE SYSTEM REVEALED A BUILD UP OF A WHITE PRECIPITATE AROUND THE PUMP 2 AND WD (WASH DISPLACEMENT) PORTS ON THE VALVE MANIFOLD. THE FSE REPLACED THE VALVE MANIFOLD AND RAN QC. THE INSTRUMENT HAS BEEN REPAIRED. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

DISCORDANT TROPONIN (TNI) RESULTS WERE OBTAINED ON THE ADVIA CENTAUR CP INSTRUMENT. THE LABORATORY RETESTED THE SAMPLES AND CORRECTED REPORTS WERE ISSUED. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES OR PATIENT INTERVENTION DUE TO THE DISCORDANT TROPONIN RESULTS.

Description of Event or Problem · 1

DISCORDANT TROPONIN (TNI) RESULTS WERE OBTAINED ON THE ADVIA CENTAUR CP INSTRUMENT. THE LABORATORY RETESTED THE SAMPLES AND CORRECTED REPORTS WERE ISSUED. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES OR PATIENT INTERVENTION DUE TO THE DISCORDANT TROPONIN RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVIA CENTAUR CP IMMUNOASSAY ANALYZER MMI SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR CP N/A

Patients

Seq Age Sex Outcome Treatment
1