FDA Adverse Event Malfunction Summary report: N

PURSTRING

MDR report key: 21228979 · Received January 24, 2025

Report

Report Number
2647580-2025-00305
Event Type
Malfunction
Date Received
January 24, 2025
Date of Event
January 8, 2025
Report Date
April 3, 2025
Manufacturer
US SURGICAL PUERTO RICO
Product Code
GDJ
UDI-DI
10884521051430
PMA / PMN Number
K901107
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D10 CONCOMITANT PRODUCT: 020242, 020242 DISP PURSE STRING 65 X3, (LOT # P2J0129). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: G3, H3, H6 H3. EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THAT THE PUSHERS ARE FULLY DEPLOYED. CARTRIDGES ARE PROPERLY SEATED AFTER FIRING. NO VISUAL ABNORMALITIES. IT WAS REPORTED THAT THE INSTRUMENT DID NOT FIRE. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE ESTABLISHED FROM THE INFORMATION AVAILABLE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING THE DEVICE, THE INSTRUMENT DID NOT FIRE. THE SAME ISSUE OCCURRED WITH THE SECOND DEVICE. THE DEVICE WAS REPLACED WITH A NEW ONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
742279 PURSTRING CLAMP, SURGICAL, GENERAL & PLASTIC SURGERY GDJ US SURGICAL PUERTO RICO 020242 P2J0129 10884521051430

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown SEE H11