FDA Adverse Event Malfunction Summary report: N

HMS PLUS INSTRUMENT

MDR report key: 21228974 · Received January 24, 2025

Report

Report Number
2184009-2025-00066
Event Type
Malfunction
Date Received
January 24, 2025
Date of Event
January 22, 2025
Report Date
March 11, 2025
Manufacturer
PERFUSION SYSTEMS
Product Code
JOX
UDI-DI
00763000158002
PMA / PMN Number
K101271
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CORRECTION H6.4 EVAL CODE CONCLUSION (FDC/ANNEX D): THIS FIELD WAS UPDATED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION SUMMARY: THE REPORTED ABNORMAL CONTROL FAIL ISSUE WAS VERIFIED DURING SERVICE. SERVICE TECHNICIAN OBSERVED ABNORMAL CONTROL WOULD PASS AND FAIL INTERMITTENTLY. INSTRUMENT PASSED ALL TEST WHEN RUNNING ELECTRONIC CONTROL. FLAG HEIGHT WAS ON THE LOW SIDE OF THE SPEC FOR CHANNEL 5 AND 6. SERVICE TECHNICIAN RAN DIAG 760, FLAG DROP TIMES WITH ALL CHANNELS WERE MEASURING IN THE HIGH 50 AND 39. ADJUSTED FLAG DROP TIMES TO BRING ALL CHANNELS UP TO 60 AND 61. RAN MULTIPLE ABNORMAL WET CONTROLS AND ALL TEST PASSED. PREVENTIVE MAINTENANCE WAS PERFORMED AS PER SPECIFICATION. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION B5. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE LOT NUMBER OF THE CONTROL CARTRIDGES IS 0230007362. BOTH LIQUID OR ELECTRONIC (ACTTRAC OR HEPTRAC) CONTROLS WERE USED. QUALITY CONTROL IS PERFORMED FOR THIS UNIT ONCE EVERY 7 DAYS. NO ERROR CODE WAS ASSOCIATED WITH THIS ISSUE. CONTROL VALUES WERE OBTAINED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A HMS PLUS INSTRUMENT, IT WAS REPORTED THAT THE UNIT WAS NOT PASSING ABNORMAL CONTROLS. THE INSTRUMENT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
742274 HMS PLUS INSTRUMENT ANALYZER, HEPARIN, AUTOMATED JOX PERFUSION SYSTEMS 30514 00763000158002

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown