FDA Adverse Event Malfunction Summary report: N

ETS FLEX ARTICNG

MDR report key: 2122854 · Received June 14, 2011

Report

Report Number
3005075853-2011-02398
Event Type
Malfunction
Date Received
June 14, 2011
Date of Event
March 31, 2011
Report Date
April 1, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. FIRING TRIGGER TEETH THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AS THE FIRING TRIGGER WAS JAMMED HALFWAY. A RELOAD WAS LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED PARTIALLY FIRED AND WITH CARTRIDGE LOCKOUT SPRING DAMAGED. THE DAMAGE TO THE RELOAD LOCKOUT SPRING IS CONSISTENT WITH DAMAGE OBSERVED WHEN THE FIRING CYCLE IS STARTED, INTERRUPTED, RELEASED, AND RESTARTED. WHEN FIRING THE DEVICE MAKE SURE THAT THE FIRING STROKE IS COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR MORE INFORMATION. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A INTRAPERITONEAL HYPERTHERMIC PERFUSION SPLENECTOMY PROCEDURE THE STAPLER JAMMED AND WOULD NOT FIRE WHILE THE SURGEON WAS PERFORMING THE SPLENECTOMY. THERE WAS NO ADDITIONAL INFORMATION AVAILABLE AT THIS TIME. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETS FLEX ARTICNG STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK

Patients

Seq Age Sex Outcome Treatment
1