FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2122623 · Received June 14, 2011

Report

Report Number
6000001-2011-07621
Event Type
Malfunction
Date Received
June 14, 2011
Date of Event
May 26, 2011
Report Date
May 27, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL NARRATIVE: THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: ONE SAMPLE WAS RECEIVED BY BAXTER FOR EVALUATION. VISUAL EXAMINATION OF THE UNIT DID CONFIRM THE LEAK. THIS IS A SINGLE USE DEVICE AND WILL NOT BE REPAIRED. NO OTHER OBSERVATION WAS FOUND ON THE UNITS. THE ROOT CAUSE WAS DUE TO THE STRESS FROM THE PACKAGING DESIGN. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT ONE (1) INTERMATE WAS LEAKING "AT CUFF END AT BASE". THE DEVICE WAS FILLED WITH VANCOMYCIN. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE

Patients

Seq Age Sex Outcome Treatment
1 VANCOMYCIN