FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 2122612 · Received June 14, 2011

Report

Report Number
2649622-2011-07891
Event Type
Injury
Date Received
June 14, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED IN-PART ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): INNER INSULATION BREACHED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE INNER AND OUTER INSULATION HAD COSMETIC METAL ION OXIDATION, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD WAS STRETCHED; FULL LEAD IN SEGMENTS RETURNED AND ANALYZED.

Description of Event or Problem · 1

THE ATRIAL LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. FOLLOW-UP WITH THE PHYSICIAN'S OFFICE DETERMINED THAT BOTH THE ATRIAL AND RIGHT VENTRICULAR LEADS WERE BROKEN AND HAD BEEN REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4068 ASKU

Patients

Seq Age Sex Outcome Treatment
1 32 YR Hospitalization| R (B)(4) IMPLANTABLE PULSE GENERATOR