FDA Adverse Event Injury Summary report: N

MAXIMO VR

MDR report key: 2122573 · Received June 13, 2011

Report

Report Number
6000094-2011-00701
Event Type
Injury
Date Received
June 13, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
P980016/S37
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. THIS INFORMATION IS BASED ENTIRELY ON JOURNAL LITERATURE. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS LIMITED DUE TO CONFIDENTIALITY CONCERNS. REFERENCED ARTICLE: "INAPPROPRIATE SHOCK DELIVERY BY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR DUE TO ELECTRICAL LEAK FROM BATHROOM TAP." CLIN. RES. CARDIOL. NOVEMBER 1;99(11):761-763. SERIAL NUMBERS HAVE BEEN ADDED.

Description of Event or Problem · 1

A JOURNAL ARTICLE WAS REVIEWED THAT CONTAINED INFORMATION REGARDING THIS DEVICE AND LEAD. IT WAS REPORTED THAT THE PATIENT EXPERIENCED INAPPROPRIATE SHOCKS. DEVICE INTERROGATION REVEALED TYPICAL EMI. THE PATIENT WAS PREPARING TO WASH HIS HANDS IN THE BATHROOM AND TOUCHED THE TAP WITH HIS WET LEFT HAND SHORTLY BEFORE THE SHOCK. THE PATIENT REMAINED ASYMPTOMATIC BEFORE AND AFTER THE SHOCK. THE EGM REVEALED NOISE. AN ELECTRICIAN INSPECTED THE BATHROOM TAP AND A "SMALL AMOUNT OF CURRENT LEAK WAS DETECTED." THE ELECTRICIAN FOUND THE LEAK ORIGINATED FROM AN "INAPPROPRIATELY EARTHED BATHROOM SOCKET." THE SOCKET WAS REGROUNDED. IT APPEARS THE DEVICE AND LEAD ARE STILL IN USE. FURTHER FOLLOW-UP DID NOT YIELD ANY ADDITIONAL RELEVANT INFORMATION REGARDING THIS EVENT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAXIMO VR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. 7232CX ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| L| R