INFANT BIAS FLOW BREATHING CIRCUIT
Report
- Report Number
- 9611451-2011-00352
- Event Type
- Malfunction
- Date Received
- June 13, 2011
- Date of Event
- May 17, 2011
- Report Date
- May 19, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- REFER TO H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE SWIVEL Y-PIECE OF THE RT125 INFANT BREATHING CIRCUIT WAS RETURNED TO FPH (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED FOR CRACK AND PRESSURE TESTED FOR LEAKS. RESULTS: VISUAL INSPECTION REVEALED THAT THE RETURNED SWIVEL Y-PIECE HAD A CRACK ON THE PORTS WHERE THE BREATHING CIRCUITS ARE CONNECTED. THE PRESSURE TEST RESULT WAS WITHIN THE REQUIRED SPECIFICATION. A LOT CHECK WAS NOT PERFORMED AS NO LOT INFORMATION HAD BEEN PROVIDED. CONCLUSION: ALL BREATHING CIRCUITS, INCLUDING THE SWIVELS, ARE VISUALLY INSPECTED AND PRESSURE TESTED BEFORE THEY ARE ALLOWED TO LEAVE THE PRODUCT LINE. THE AFFECTED SWIVEL Y-PIECE WOULD HAVE MET THE REQUIRED SPECIFICATION AT THE TIME OF PRODUCTION. THIS SUGGESTS THE CRACK DEVELOPED POST-PRODUCTION, POSSIBLY THE BREATHING CIRCUITS WERE CONNECTED TO OR REMOVED FROM THE SWIVEL FORCEFULLY. THE RT125 USER INSTRUCTIONS THAT ACCOMPANY THE DEVICE STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE". "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT". "SET APPROPRIATE VENTILATOR ALARMS". "THIS PRODUCT IS INTENDED TO BE USED FOR A MAXIMUM OF 7 DAYS". (B)(4).
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL (FPH) THAT THE SWIVEL Y-PIECE OF AN RT125 INFANT BIAS FLOW BREATHING CIRCUIT HAD CRACKED AFTER (B)(6) OF USE. IT WAS ALSO REPORTED THAT AIR LEAKED FROM THE SWIVEL. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT BIAS FLOW BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | RT125 | NOT PROVIDED |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |