FDA Adverse Event Malfunction Summary report: N

INFANT BIAS FLOW BREATHING CIRCUIT

MDR report key: 2122551 · Received June 13, 2011

Report

Report Number
9611451-2011-00352
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
May 17, 2011
Report Date
May 19, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
REFER TO H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE SWIVEL Y-PIECE OF THE RT125 INFANT BREATHING CIRCUIT WAS RETURNED TO FPH (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED FOR CRACK AND PRESSURE TESTED FOR LEAKS. RESULTS: VISUAL INSPECTION REVEALED THAT THE RETURNED SWIVEL Y-PIECE HAD A CRACK ON THE PORTS WHERE THE BREATHING CIRCUITS ARE CONNECTED. THE PRESSURE TEST RESULT WAS WITHIN THE REQUIRED SPECIFICATION. A LOT CHECK WAS NOT PERFORMED AS NO LOT INFORMATION HAD BEEN PROVIDED. CONCLUSION: ALL BREATHING CIRCUITS, INCLUDING THE SWIVELS, ARE VISUALLY INSPECTED AND PRESSURE TESTED BEFORE THEY ARE ALLOWED TO LEAVE THE PRODUCT LINE. THE AFFECTED SWIVEL Y-PIECE WOULD HAVE MET THE REQUIRED SPECIFICATION AT THE TIME OF PRODUCTION. THIS SUGGESTS THE CRACK DEVELOPED POST-PRODUCTION, POSSIBLY THE BREATHING CIRCUITS WERE CONNECTED TO OR REMOVED FROM THE SWIVEL FORCEFULLY. THE RT125 USER INSTRUCTIONS THAT ACCOMPANY THE DEVICE STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE". "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT". "SET APPROPRIATE VENTILATOR ALARMS". "THIS PRODUCT IS INTENDED TO BE USED FOR A MAXIMUM OF 7 DAYS". (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL (FPH) THAT THE SWIVEL Y-PIECE OF AN RT125 INFANT BIAS FLOW BREATHING CIRCUIT HAD CRACKED AFTER (B)(6) OF USE. IT WAS ALSO REPORTED THAT AIR LEAKED FROM THE SWIVEL. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT BIAS FLOW BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED RT125 NOT PROVIDED

Patients

Seq Age Sex Outcome Treatment
1