FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2122549 · Received June 13, 2011

Report

Report Number
6000001-2011-07599
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
May 31, 2011
Report Date
June 1, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION OF REPEATING LINES ON DISPLAY WAS CONFIRMED BY SERVICE. THE CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE A FAULTY MAIN DISPLAY. THE MAIN DISPLAY WAS REPLACED TO FIX THE REPORTED CONDITION. ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE HAS BEEN ESCALATED TO CAPA. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORAMTION: THIS INVOLVED A REMEDIATED COLLEAGUE 2006 INFUSION PUMP WITH USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH REPEATING LINES ON THE DISPLAY. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS CONDITION WAS FOUND UPON POWER UP. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVED AND THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS UNKNOWN AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1