SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-07865
- Event Type
- Malfunction
- Date Received
- June 13, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. THE HELIX WAS DISENGAGED FROM THE HELICAL CHANNEL. THE DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED). THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE HELIX/LOBE MECHANISM (SLEEVE HEAD). THERE WAS A TIP SEAL OBSERVATION.
THE LEAD WAS RETURNED TO THE MANUFACTURER AFTER THE HELIX WOULD NOT EXTEND OR RETRACT DURING AN IMPLANT ATTEMPT. THE LEAD SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Other |