FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2122501 · Received June 13, 2011

Report

Report Number
2649622-2011-07865
Event Type
Malfunction
Date Received
June 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. THE HELIX WAS DISENGAGED FROM THE HELICAL CHANNEL. THE DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED). THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE HELIX/LOBE MECHANISM (SLEEVE HEAD). THERE WAS A TIP SEAL OBSERVATION.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER AFTER THE HELIX WOULD NOT EXTEND OR RETRACT DURING AN IMPLANT ATTEMPT. THE LEAD SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 73 YR Other