ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-04110
- Event Type
- Malfunction
- Date Received
- June 13, 2011
- Date of Event
- May 13, 2011
- Report Date
- May 17, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(6). THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE DISTRIBUTOR REPORTED THAT THE PATIENT NEEDED TO PRESS THE BUTTONS SEVERAL TIMES TO ACHIEVE DESIRED PUMP FUNCTIONS. NO OTHER TROUBLESHOOTING DETAIL WAS PROVIDED. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |