CONCERTO CRT-D DR
Report
- Report Number
- 2647346-2011-00795
- Event Type
- Injury
- Date Received
- June 13, 2011
- Date of Event
- January 18, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE ICD DEVICE AND HAVE ANALYZED THIS DATA. HIGH IMPEDANCE WAS NOTED. OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE PATIENT ALERT IS OBSERVED ON (B)(6) 2011. INTERFERENCE/NOISE WAS NOTED. HIGH VENTRICULAR SHORT INTERVAL COUNT (SIC) IS OBSERVED WITH 510 COUNTS OCCURRING BETWEEN THE (B)(6) 2011 AND THE (B)(6) 2011. ZERO COUNTS ARE RECORDED BETWEEN THE (B)(6) AND THE (B)(6) 2011. HIGH SIC CAN INDICATE THE PRESENCE OF NOISE OR INTERMITTENCY IN THE SYSTEM.
IT WAS REPORTED THAT THE LEAD HAD AN APPARENT FRACTURE OR A BREACH. THERE WAS OVERSENSING (NOISE), NON-SUSTAINED TACHYCARDIA AND VENTRICULAR TACHYCARDIA EPISODES BUT NOT THERAPY. IT WAS ALSO REPORTED THERE WAS INHIBITION OF PACING AND THE LEAD HAD OVER AND UNDER SENSING. THE PACE/SENSE PORTION OF THE LEAD WAS CAPPED AND REPLACED BY A NEW PACE/SENSE LEAD. THE HIGH VOLTAGE PART OF THE LEAD IS STILL IN USE. IT WAS FURTHER REPORTED THERE WAS ASYSTOLE. IT WAS ALSO REPORTED THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) MALFUNCTIONED. THE ICD DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCERTO CRT-D DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | C154DWK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD |