FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2122490 · Received June 13, 2011

Report

Report Number
2649622-2011-07859
Event Type
Injury
Date Received
June 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE PROXIMAL CONDUCTOR HAD FRACTURED. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION WAS TORN, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX MECHANISM, AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

THE LEAD WAS REPLACED BECAUSE THE PHYSICIAN WANTED TO REDUCE THE AMOUNT OF HARDWARE IN THE PATIENT. THE LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| R 4076 ATRIAL IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR