FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2122469 · Received June 13, 2011

Report

Report Number
2531779-2011-04106
Event Type
Malfunction
Date Received
June 13, 2011
Report Date
May 16, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED ON (B)(6) 2011. ALL KEYPAD BUTTONS WERE INTERMITTENTLY ACTIVATING PUMP FUNCTIONS WHEN PRESSED. ADHESIVE WAS FOUND UNDER BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2011, THE LAY-USER/PATIENT CONTACTED ANIMAS ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT CLAIMED THAT THE PUMP WAS UNRESPONSIVE TO BUTTON PRESSES. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 1250 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 19 YR