FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2122456 · Received June 13, 2011

Report

Report Number
2649622-2011-07851
Event Type
Injury
Date Received
June 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYSIS REVEALED NO ANOMALIES. IT WAS NOTED THAT THE DISTAL CONDUCTOR WAS FRACTURED (OVERSTRESS), THE INNER TUBING WAS KINKED/BUCKLED, THERE WAS BLOOD IN/ON HELIX/LOBE MECHANISM AND MECHANISM (SLEEVE HEAD) AND THERE WAS APPARENT IMPLANT DAMAGE.

Description of Event or Problem · 1

THE LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R