FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2122449 · Received June 13, 2011

Report

Report Number
1423500-2011-07551
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
May 19, 2011
Report Date
May 19, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PER THE CUSTOMER THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A LOW DRAIN VOLUME ALARM WAS NOT CONFIRMED. THE CAUSE WAS AIR IN PATIENT LINE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION FOR THE LOW DRAIN VOLUME ALARMS IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).

Description of Event or Problem · 1

DURING TROUBLESHOOTING OF LOW DRAIN VOLUME (LDV) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING INITIAL DRAIN, THE HOME PATIENT (HP) REVEALED THAT THERE WERE LOTS OF AIR IN THE PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED TO THE HP TO CHECK THE LINE FOR AIR BUBBLES AFTER PRIME, BUT BEFORE CONNECTING. THE TSR THEN ASSISTED THE HP IN ENDING THERAPY AND ADVISED TO START OVER WITH NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. DURING A FOLLOW UP WITH HP REGARDING THE AIR IN LINE, IT WAS REVEALED THAT THE ISSUE WAS RESOLVED BY STARTING OVER USING ALL NEW SUPPLIES. THE HP DID NOT KNOW THE CAUSE OF THE ALARM AND THAT IT WAS AN ISOLATED EVENT. PER HP, THERE WERE NO LOOSE CONNECTIONS, DISCONNECTIONS OR DEFECTS ON THE SUPPLIES. THE HP STATED THAT SHE IS DOING FINE AND CONTINUING THERAPY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 73 YR HOMECHOICE