FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2122434 · Received June 13, 2011

Report

Report Number
6000144-2011-02161
Event Type
Injury
Date Received
June 13, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Description of Event or Problem · 1

THE DEVICE WAS REPLACED DUE TO ERI (ELECTIVE REPLACEMENT INDICATOR). THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D224TRK ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R 6949 IMPLANTABLE TACHY LEAD| 4537 COMPETITOR IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD