FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2122381 · Received June 13, 2011

Report

Report Number
2531779-2011-04098
Event Type
Malfunction
Date Received
June 13, 2011
Report Date
May 15, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE OK/DOWN KEYPAD BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DOWN ARROW KEYPAD BUTTON HAD TO BE PRESSED MULTIPLE TIMES IN ORDER TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION OTP GLUCOSE MGMT SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 34 YR