FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2122381
·
Received June 13, 2011
Report
- Report Number
- 2531779-2011-04098
- Event Type
- Malfunction
- Date Received
- June 13, 2011
- Report Date
- May 15, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE OK/DOWN KEYPAD BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DOWN ARROW KEYPAD BUTTON HAD TO BE PRESSED MULTIPLE TIMES IN ORDER TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | OTP GLUCOSE MGMT SYSTEM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |