FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2122380 · Received June 13, 2011

Report

Report Number
6000001-2011-07585
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
May 20, 2011
Report Date
May 20, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION RESULTS: THE REPORTED CONDITION OF BATTERY DEPLETED SET ALARM WAS DISCOVERED AND WAS CONFIRMED DURING PRODUCT EVALUATION IN THE PUMP'S EVENT HISTORY. THIS CONDITION WAS CAUSED BY A DEPLETED BATTERY ALARM. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Description of Event or Problem · 1

DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY (B)(4) PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED DEPLETED BATTERY ALARM, WHICH INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.08.92.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1