FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2122365 · Received June 13, 2011

Report

Report Number
6000001-2011-07583
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
May 19, 2011
Report Date
May 19, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW UP MEDWATCH WILL BE SUBMITTED UPON COMPLETION OF BAXTER'S INVESTIGATION. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH DOWNSTREAM OCCLUSIONS WAS CONFIRMED DURING PRODUCT EVALUATION. DELIVERY ACCURACY TESTS WERE PERFORMED AND FOUND THE PUMP TO BE DELIVERING WITHIN SPECIFICATIONS. THEREFORE, NO ASSIGNABLE CAUSE COULD BE PROVIDED FOR THE REPORTED CONDITION AND NO REPAIR WAS NECESSARY. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED TO BAXTER (B)(4) TECHNICAL SERVICES ONE COLLEAGUE INFUSION WITH DOWNSTREAM OCCLUSIONS. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED HOWEVER, THE CONDITION OCCURRED IN BIOMED SERVICE DEPARTMENT. THERE IS NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THIS EVENT. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THIS DEVICE. THE SOFTWARE VERSION FOR THIS DEVICE IS CURRENTLY NOT KNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1