FDA Adverse Event Malfunction Summary report: N

TREK CORONARY DILATATION CATHETER

MDR report key: 2122359 · Received June 13, 2011

Report

Report Number
2024168-2011-04148
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
May 19, 2011
Report Date
May 19, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K103110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. FACTORS THAT CAN CONTRIBUTE TO BALLOON RUPTURES INCLUDE, BUT ARE NOT LIMITED TO: BALLOON DAMAGE DURING MANUFACTURING, WEAK MATERIALS, EXCESSIVELY APPLIED PRESSURE, OR INTERACTION WITH ACCESSORY DEVICES, PATIENT ANATOMY, AND/OR LESION CALCIFICATION OR TORTUOSITY. TO ENSURE THIS IS NOT A RESULT OF MANUFACTURING, ALL BALLOON CATHETERS ARE 100% VISUALLY INSPECTED AND LEAK TESTED PRIOR TO PACKAGING. ADDITIONALLY, A SAMPLING OF UNITS IS DESTRUCTIVELY TESTED TO VERIFY THE RATED BURST PRESSURE (RBP). RETURN OF THE RX TREK CATHETER USED IN THE PROCEDURE MAY HAVE AIDED IN THE INVESTIGATION AND DETERMINATION OF CAUSE. THERE WAS NO REPORT OF ANY LEAK IN THE CATHETER NOTED DURING PREPARATION FOR USE, WHICH WOULD SUGGEST THAT THE BALLOON WAS NOT DAMAGED PRIOR TO USE. IT IS POSSIBLE THAT THE BALLOON MATERIAL WAS DAMAGED (SCRATCHED) DURING INTERACTIONS WITH THE HEAVILY CALCIFIED LESION, SUCH THAT THE BALLOON RUPTURED UPON THE FIRST INFLATION AT 10 ATMOSPHERES WHICH IS BELOW THE RBP. A REVIEW OF THE PRODUCT MANUFACTURING RECORDS DID NOT REVEAL ANY NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT AND ALL LOT RELEASE TESTING MET MANUFACTURING CRITERIA. ADDITIONALLY, A QUERY OF THE COMPLAINT-HANDLING DATABASE WAS PERFORMED AND THERE HAVE BEEN NO OTHER INCIDENTS REPORTED FOR BALLOON RUPTURES FOR THIS LOT. THERE IS NO INDICATION OF A LOT SPECIFIC PRODUCT QUALITY DEFICIENCY. A DEFINITIVE CAUSE FOR THE REPORTED BALLOON RUPTURE CANNOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRE-DILATATION WAS PERFORMED WITH THE TREK BALLOON CATHETER DURING A MID RIGHT CORONARY ARTERY PROCEDURE. UPON FIRST INFLATION, THE TREK BALLOON RUPTURED AFTER 15 SECONDS, AT 10 ATMOSPHERES. VESSEL AND LESION SITE WERE CHARACTERIZED AS BEING MODERATELY TORTUOUS, HEAVILY CALCIFIED, CONCENTRIC, DE NOVO AND 99% STENOSED. A NON-ABBOTT BALLOON CATHETER WAS USED TO CONTINUE WITH THE PRE-DILATATION PRIOR TO DEPLOYMENT OF A XIENCE V STENT TO COMPLETE THE PROCEDURE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TREK CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 0112262

Patients

Seq Age Sex Outcome Treatment
1 GW: BALANCE, FILDER FCGC: BRITE T