COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING 44MM
Report
- Report Number
- 1825034-2011-00502
- Event Type
- Injury
- Date Received
- June 13, 2011
- Date of Event
- May 19, 2011
- Report Date
- May 19, 2011
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWS
- PMA / PMN Number
- K080642
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER TEN STATES, "FATIGUE FRACTURE OF COMPONENT CAN OCCUR AS A RESULT OF LOSS OF FIXATION, STRENUOUS ACTIVITY, MALALIGNMENT, TRAUMA, NON-UNION, OR EXCESSIVE WEIGHT." ALTHOUGH POST FRACTURE DAMAGE HAS OBFUSCATED SOME ARTIFACTS, THE FRACTURE APPEARS TO HAVE STARTED IN FATIGUE AND FAILED IN A FAST-FRACTURE. THE USER FACILITY WAS NOTIFIED OF THE EVENT ON (B)(6), 2011. TO DATE, A RESPONSE HAS NOT BEEN RECEIVED FROM THE USER FACILITY. IN THE EVENT THAT THE USER FACILITY SUBMITS A MEDWATCH REPORT, BIOMET WILL FORWARD A COPY TO THE FDA. THIS REPORT SUBMITTED (B)(4), 2011.
IT WAS REPORTED THAT PATIENT UNDERWENT REVERSE SHOULDER ARTHROPLASTY ON (B)(6), 2009. SUBSEQUENTLY, THE PATIENT SUSTAINED A FALL AND HEARD A SNAP FROM THE SHOULDER. THE PATIENT WAS REVISED ON (B)(6), 2011, DUE TO FRACTURE OF THE HUMERAL TRAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING 44MM | PROSTHESIS, SHOULDER | KWS | BIOMET ORTHOPEDICS | N/A | 848310 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |