FDA Adverse Event Injury Summary report: N

COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING 44MM

MDR report key: 2122338 · Received June 13, 2011

Report

Report Number
1825034-2011-00502
Event Type
Injury
Date Received
June 13, 2011
Date of Event
May 19, 2011
Report Date
May 19, 2011
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWS
PMA / PMN Number
K080642
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER TEN STATES, "FATIGUE FRACTURE OF COMPONENT CAN OCCUR AS A RESULT OF LOSS OF FIXATION, STRENUOUS ACTIVITY, MALALIGNMENT, TRAUMA, NON-UNION, OR EXCESSIVE WEIGHT." ALTHOUGH POST FRACTURE DAMAGE HAS OBFUSCATED SOME ARTIFACTS, THE FRACTURE APPEARS TO HAVE STARTED IN FATIGUE AND FAILED IN A FAST-FRACTURE. THE USER FACILITY WAS NOTIFIED OF THE EVENT ON (B)(6), 2011. TO DATE, A RESPONSE HAS NOT BEEN RECEIVED FROM THE USER FACILITY. IN THE EVENT THAT THE USER FACILITY SUBMITS A MEDWATCH REPORT, BIOMET WILL FORWARD A COPY TO THE FDA. THIS REPORT SUBMITTED (B)(4), 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT UNDERWENT REVERSE SHOULDER ARTHROPLASTY ON (B)(6), 2009. SUBSEQUENTLY, THE PATIENT SUSTAINED A FALL AND HEARD A SNAP FROM THE SHOULDER. THE PATIENT WAS REVISED ON (B)(6), 2011, DUE TO FRACTURE OF THE HUMERAL TRAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING 44MM PROSTHESIS, SHOULDER KWS BIOMET ORTHOPEDICS N/A 848310

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R