ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-04089
- Event Type
- Malfunction
- Date Received
- June 13, 2011
- Report Date
- May 17, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. THE BUTTONS BECOME STUCK WHEN PRESSED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PATIENT REPORTED THAT THE DOWN ARROW BUTTON DOES NOT RESPOND AND SOMETIMES STICKS. THE PATIENT REPORTEDLY WORE THE PUMP ON THE OUTSIDE OF HER CLOTHING AND CLEANED THE PUMP WITH (B)(6). (B)(4) ADVISED THE PATIENT NOT TO CLEAN THE PUMP WITH (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |