SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-07841
- Event Type
- Malfunction
- Date Received
- June 13, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED IN SEGMENTS. ANALYSIS FOUND THAT THE DISTAL CONDUCTOR WAS FRACTURED, THAT THE INNER TUBING WAS KINKED/BUCKLED, THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THAT THE HELIX/LOBE WAS DISTORTED/BENT AND THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. THE ANALYST ALSO OBSERVED THAT THE LEAD APPEARED TO HAVE BEEN IMPLANTED WITH A BEND IN IT AT THE FRACTURE SITE.
THE RIGHT VENTRICULAR LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Hospitalization| R |