HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-07524
- Event Type
- Malfunction
- Date Received
- June 13, 2011
- Date of Event
- May 25, 2011
- Report Date
- May 25, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE IS NOT AVAILABLE AND THE LOT NUMBER IS UNKNOWN. SINCE THE LOT NUMBER IS UNKNOWN, NO BATCH REVIEW WILL BE PERFORMED. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED.
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED. THE CAUSE OF THE SYSTEM ERROR 2240 WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE POTENTIAL USE ERROR IN THIS COMPLAINT. BAXTER FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT A SYSTEM ERROR (SE) 2240 WHICH OCCURRED ON HOME CHOICE (HC) DURING USE DURING THERAPY. THE BAXTER TECHNICAL REPRESENTATIVE (TSR) EXPLAINED THE SE AND HAD THE HOME PATIENT (HP) CYCLE POWER OFF AND ON AND GOT SE 2367. THE TSR EXPLAINED TO THE HP THE PROPER PROCEDURES FOR WHEN SHE WANTS TO DISCONNECT. THE TSR HAD THE HP CYCLE POWER OFF AND ON AGAIN. THE SE DID NOT REPEAT. THE HC WAS AT "PRESS GO TO START" . THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. (B)(4) CONTACTED THE HP ON (B)(6) 2011 REGARDING THE SYSTEM ERROR 2240 ALARM. PER THE HP, SHE DISCONNECTED TO GO TO THE KITCHEN AND WHEN SHE GOT BACK SHE DIDN'T THINK ABOUT RECONNECTING. THE HC ALARMED AND THEN THE HP RECONNECTED. THE WRITER ADVISED THE HP THAT IT IS NOT RECOMMENDED THAT THE HP RECONNECT AFTER THE HC HAS MOVED TO ANOTHER CYCLE AS THE ALARM MEANS THERE IS AIR IN THE SET. THE HP UNDERSTOOD THE ALARM. THE HP STATED SHE WAS IN THE FOURTH CYCLE, SO SHE DID NOT START OVER WITH NEW SUPPLIES AND SHE DID CONTACT HER PD RN ABOUT THE ALARM. THE HP STATED SHE RESUMED THERAPY THE FOLLOWING NIGHT ON THE CYCLER. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.