FDA Adverse Event Malfunction Summary report: N

EXCEL 36KHZ MICROTIP PLUS BOX4

MDR report key: 2122090 · Received May 24, 2011

Report

Report Number
1222895-2011-00010
Event Type
Malfunction
Date Received
May 24, 2011
Date of Event
March 3, 2011
Report Date
May 24, 2011
Manufacturer
INTEGRA, BURLINGTON
Product Code
LFL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED CUSA TIP WAS EXAMINED AND IT WAS DETERMINED THE TIP WAS BLOCKED WITH AN UNK SUBSTANCE. IF THE BLOCKAGE OCCURRED DURING THE PROCEDURE THE TIP WOULD OVERHEAT AND CAUSE THE TIP TO BREAK. INTEGRA CONSIDERS THIS COMPLAINT INVESTIGATION CLOSED. FUTURE INCIDENTS OF THIS NATURE WILL BE DOCUMENTED FOR RECURRENCE AND TRENDING PURPOSES.

Description of Event or Problem · 1

A CUSA MICROTIP PLUS WAS INVOLVED IN AN INCIDENT AND WAS DESCRIBED AS FOLLOWS: THE SURGEON WAS TREATING A CRANIOPHARYNGIOMA (HARD TUMOR), ON A CHILD WHILE UTILIZING A CUSA MICROTIP PLUS (B)(4). THE SURGERY WAS COMPLETED WHEN THE TIP BROKE. THERE WAS NO INJURY OR DELAY IN SURGERY AS A RESULT OF THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXCEL 36KHZ MICROTIP PLUS BOX4 ULTRASONIC SURGICAL PRODUCTS LFL INTEGRA, BURLINGTON

Patients

Seq Age Sex Outcome Treatment
1