FDA Adverse Event
Malfunction
Summary report: N
EXCEL 36KHZ MICROTIP PLUS BOX4
MDR report key: 2122090
·
Received May 24, 2011
Report
- Report Number
- 1222895-2011-00010
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Date of Event
- March 3, 2011
- Report Date
- May 24, 2011
- Manufacturer
- INTEGRA, BURLINGTON
- Product Code
- LFL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE RETURNED CUSA TIP WAS EXAMINED AND IT WAS DETERMINED THE TIP WAS BLOCKED WITH AN UNK SUBSTANCE. IF THE BLOCKAGE OCCURRED DURING THE PROCEDURE THE TIP WOULD OVERHEAT AND CAUSE THE TIP TO BREAK. INTEGRA CONSIDERS THIS COMPLAINT INVESTIGATION CLOSED. FUTURE INCIDENTS OF THIS NATURE WILL BE DOCUMENTED FOR RECURRENCE AND TRENDING PURPOSES.
Description of Event or Problem · 1
A CUSA MICROTIP PLUS WAS INVOLVED IN AN INCIDENT AND WAS DESCRIBED AS FOLLOWS: THE SURGEON WAS TREATING A CRANIOPHARYNGIOMA (HARD TUMOR), ON A CHILD WHILE UTILIZING A CUSA MICROTIP PLUS (B)(4). THE SURGERY WAS COMPLETED WHEN THE TIP BROKE. THERE WAS NO INJURY OR DELAY IN SURGERY AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EXCEL 36KHZ MICROTIP PLUS BOX4 | ULTRASONIC SURGICAL PRODUCTS | LFL | INTEGRA, BURLINGTON |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |