FDA Adverse Event Malfunction Summary report: N

LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL)

MDR report key: 2122089 · Received May 24, 2011

Report

Report Number
2024601-2011-00412
Event Type
Malfunction
Date Received
May 24, 2011
Date of Event
January 15, 2011
Report Date
May 2, 2011
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ALLERGAN HAS RECEIVED THE PRODUCT, HOWEVER, THE DEVICE HAS NOT BEEN IDENTIFIED NOR HAS THE ANALYSIS BEEN COMPLETED AT THIS TIME. BASED UPON THE MODEL NUMBER, SERIAL NUMBER AND IMPLANT DATE PROVIDED BY THE REPORTER, THE CONNECTOR TYPE IS ASSUMED TO BE A TAPER II. ALLERGAN HAS RECEIVED THE PRODUCT, HOWEVER, THE ANALYSIS HAS NOT BEEN COMPLETED AT THIS TIME. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING." DEVICE LABELING ADDRESSES THE FOLLOWING CONCERNS TO PERFORMING AN ADJUSTMENT: "PRIOR TO DOING AN ADJUSTMENT TO DECREASE THE STOMA, REVIEW THE PT'S CHART FOR TOTAL BAND VOLUME AND RECENT ADJUSTMENTS. IF RECENT ADJUSTMENTS HAVE NOT BEEN EFFECTIVE IN INCREASING RESTRICTION AND THE PT HAS BEEN COMPLIANT WITH NUTRITIONAL GUIDELINES, THE PT MAY HAVE A LEAKING BAND SYSTEM, OR MAY HAVE POUCH ENLARGEMENT OR ESOPHAGEAL DILATATION DUE TO STOMAL OBSTRUCTION, BAND SLIPPAGE OR OVER-RESTRICTION."

Description of Event or Problem · 1

EVENT WAS REPORTED AS "PATIENT WAS NOT FEELING RESTRICTION AFTER SEVERAL FILLS. WE ASPIRATED SEVERAL TIMES AND FOUND THAT SYSTEM WAS NOT HOLDING SALINE." THE PORT WAS EXPLANTED AND REPLACED. THE SURGEON NOTED LEAK WAS AT THE PORT HOUSING AND A "SEAL LEAK" WAS OBSERVED. THE DEVICE IS BEING RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL) LTI ALLERGAN NA 1838139

Patients

Seq Age Sex Outcome Treatment
1