LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL)
Report
- Report Number
- 2024601-2011-00412
- Event Type
- Malfunction
- Date Received
- May 24, 2011
- Date of Event
- January 15, 2011
- Report Date
- May 2, 2011
- Manufacturer
- ALLERGAN
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). ALLERGAN HAS RECEIVED THE PRODUCT, HOWEVER, THE DEVICE HAS NOT BEEN IDENTIFIED NOR HAS THE ANALYSIS BEEN COMPLETED AT THIS TIME. BASED UPON THE MODEL NUMBER, SERIAL NUMBER AND IMPLANT DATE PROVIDED BY THE REPORTER, THE CONNECTOR TYPE IS ASSUMED TO BE A TAPER II. ALLERGAN HAS RECEIVED THE PRODUCT, HOWEVER, THE ANALYSIS HAS NOT BEEN COMPLETED AT THIS TIME. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING." DEVICE LABELING ADDRESSES THE FOLLOWING CONCERNS TO PERFORMING AN ADJUSTMENT: "PRIOR TO DOING AN ADJUSTMENT TO DECREASE THE STOMA, REVIEW THE PT'S CHART FOR TOTAL BAND VOLUME AND RECENT ADJUSTMENTS. IF RECENT ADJUSTMENTS HAVE NOT BEEN EFFECTIVE IN INCREASING RESTRICTION AND THE PT HAS BEEN COMPLIANT WITH NUTRITIONAL GUIDELINES, THE PT MAY HAVE A LEAKING BAND SYSTEM, OR MAY HAVE POUCH ENLARGEMENT OR ESOPHAGEAL DILATATION DUE TO STOMAL OBSTRUCTION, BAND SLIPPAGE OR OVER-RESTRICTION."
EVENT WAS REPORTED AS "PATIENT WAS NOT FEELING RESTRICTION AFTER SEVERAL FILLS. WE ASPIRATED SEVERAL TIMES AND FOUND THAT SYSTEM WAS NOT HOLDING SALINE." THE PORT WAS EXPLANTED AND REPLACED. THE SURGEON NOTED LEAK WAS AT THE PORT HOUSING AND A "SEAL LEAK" WAS OBSERVED. THE DEVICE IS BEING RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL) | LTI | ALLERGAN | NA | 1838139 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |