CAPSUREFIX NOVUS

MDR report key: 2122021 · Received June 13, 2011

Report

Report Number
2649622-2011-07831
Event Type
Injury
Date Received
June 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED AND PER ANALYSIS FINDINGS THE DISTAL CONDUCTOR WAS FRACTURED, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

THE RIGHT ATRIAL LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. IT WAS REMOVED AND NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 81 YR Hospitalization| R 6949 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB