Received June 2, 2011
Report
- Report Number
- 1831750-2011-05433
- Event Type
- Other
- Date Received
- June 2, 2011
- Date of Event
- May 5, 2011
- Report Date
- May 5, 2011
- Manufacturer
- STRYKER MEDICAL
- Product Code
- FZM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
IT WAS REPORTED BY SVC REPORT THAT THE BRAKES ARE NOT WORKING. AN EMPLOYEE OF THE FACILITY REPAIRED IT PRIOR TO THE SVC TECH ARRIVING TO EVALUATE AND REPAIR THE STOOL. IT WAS REPORTED THAT THE USER FACILITY EMPLOYEE HAD REPLACED THE BRAKE ASSEMBLY AND PUT THE SEAT ON BACKWARDS. IT WAS FURTHER REPORTED THAT ONE OF THE DOCTORS SAT ON THE STOOL, PUSHED BACKWARDS AND FELL OFF THE CHAIR WHEN THE BRAKE DID NOT HOLD. NO INJURIES HAVE BEEN REPORTED. THERE WAS PT INVOLVEMENT; HOWEVER, THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SURGISTOOL | STOOL, OPERATING-ROOM | FZM | STRYKER MEDICAL | 0830 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |