FDA Adverse Event Malfunction Summary report: N

LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL)

MDR report key: 2121855 · Received May 20, 2011

Report

Report Number
2024601-2011-00370
Event Type
Malfunction
Date Received
May 20, 2011
Date of Event
May 5, 2010
Report Date
April 25, 2011
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TAPER II. (B)(4). THE DEVICE HAS BEEN RECEIVED BY ALLERGAN BUT THE CONNECTOR TYPE HAS NOT BEEN CONFIRMED BY THE DEVICE ANALYSIS LAB. BASED UPON THE MODEL NUMBER, SERIAL NUMBER AND IMPLANT DATE PROVIDED BY THE REPORTER, THE CONNECTOR TYPE IS ASSUMED TO BE A TAPER II. VISUAL EXAMINATION MAY DETERMINE THE CONNECTOR TYPE ASSOCIATED WITH THIS REPORT. ALLERGAN HAS RECEIVED THE PRODUCT; HOWEVER, THE DEVICE HAS NOT BEEN IDENTIFIED NOR HAS THE ANALYSIS BEEN COMPLETED AT THIS TIME. BASED UPON THE MODEL NUMBER, SERIAL NUMBER AND IMPLANT DATE PROVIDED BY THE REPORTER, THE CONNECTOR TYPE IS ASSUMED TO BE A TAPER II. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING." DEVICE LABELING ADDRESSES THE FOLLOWING CONCERNS TO PERFORMING AN ADJUSTMENT: "PRIOR TO DOING AN ADJUSTMENT TO DECREASE THE STOMA, REVIEW THE PT'S CHART FOR TOTAL BAND VOLUME AND RECENT ADJUSTMENTS. IF RECENT ADJUSTMENTS HAVE NOT BEEN EFFECTIVE IN INCREASING RESTRICTION AND THE PT HAS BEEN COMPLIANT WITH NUTRITIONAL GUIDELINES, THE PT MAY HAVE A LEAKING BAND SYSTEM, OR MAY HAVE POUCH ENLARGEMENT OR ESOPHAGEAL DILATATION DUE TO STOMAL OBSTRUCTION, BAND SLIPPAGE OR OVER-RESTRICTION."

Description of Event or Problem · 1

THE EVENT IS REPORTED AS "PORT LEAKING. FLUID MISSING FROM PORT." THE ACCESS PORT WAS REMOVED AND REPLACED. FOLLOW-UP FINDINGS: SURGEON NOTED THE LEAKING WAS AROUND THE MEMBRANE UNDER HIGH PRESSURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL) LTI ALLERGAN NA 1412771

Patients

Seq Age Sex Outcome Treatment
1 27 YR PER SURGEON - NO CONCOMITANT THERAPY KNOWN.