FDA Adverse Event Injury Summary report: N

CAPSURE SP

MDR report key: 2121808 · Received June 13, 2011

Report

Report Number
2649622-2011-07825
Event Type
Injury
Date Received
June 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S12
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4)-THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND THE INNER INSULATION WAS BREACHED AND HAD METAL INDUCED OXIDATION. THE ANALYST NOTED ALL CONDUCTORS WERE DISTORTED, THE DISTAL CONDUCTOR WAS STRETCHED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), ALL INSULATORS HAD COSMETIC HAD METAL INDUCED OXIDATION, THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING AND THE OUTER INSULATION WAS BREACHED CUT. IT WAS NOTED THAT THE LEAD WAS STRETCHED.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4024 ASKU

Patients

Seq Age Sex Outcome Treatment
1 82 YR Hospitalization| R 4058M IMPLANTABLE PACING LEAD| 7950B IMPLANTABLE PULSE GENERATOR