FDA Adverse Event
Malfunction
Summary report: N
HUDSON CONCHATHERM NEPTUNE HEATED HUMIDIFIER
MDR report key: 2121721
·
Received May 19, 2011
Report
- Report Number
- 3003898360-2011-00253
- Event Type
- Malfunction
- Date Received
- May 19, 2011
- Date of Event
- May 5, 2011
- Report Date
- May 6, 2011
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- BTT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE SAMPLE WAS NOT RECEIVED BY THE MFR AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN COMPLETION OF INVESTIGATION.
Description of Event or Problem · 1
THE EVENT IS REPORTED AS: COMPLAINT ALLEGES THAT THE HEATER DID NOT ALARM WHEN THE TEMP PROBE CAME OUT OF THE CIRCUIT. THE NEONATOLOGIST STATED THAT THE HEATER GOT REALLY HOT AND THE ODOR OF BURNING PLASTIC COULD BE SMELLED. THE HEATER WAS BEING USED WITH A (B)(4) COMFORT FLO HUMIDIFICATION SYSTEM AND WAS ENCLOSED IN AN ISOLETTE. THE TEMP PROBE WAS POSITIONED AT THE PT END. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HUDSON CONCHATHERM NEPTUNE HEATED HUMIDIFIER | HEATED HUMIDIFIER | BTT | TELEFLEX MEDICAL | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | COMFORT FLO HUMIDIFICATION SYSTEM, (B)(4) |