FDA Adverse Event Malfunction Summary report: N

ADJUSTABLE PIN COLLET

MDR report key: 2121541 · Received May 6, 2011

Report

Report Number
1811755-2011-01630
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
March 30, 2011
Report Date
April 1, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
GET
PMA / PMN Number
K943323
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PIN WAS STICKING IN THE DEVICE DURING ROUTINE MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADJUSTABLE PIN COLLET GET STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK