FDA Adverse Event
Malfunction
Summary report: N
ADJ PIN COLLET 2.0-3.2MM
MDR report key: 2121536
·
Received May 6, 2011
Report
- Report Number
- 1811755-2011-01635
- Event Type
- Malfunction
- Date Received
- May 6, 2011
- Date of Event
- April 20, 2011
- Report Date
- April 20, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO NOT RETAIN A PIN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT RETAIN A PIN DURING ROUTINE MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADJ PIN COLLET 2.0-3.2MM | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |