FDA Adverse Event
Malfunction
Summary report: N
ADJ PIN COLLET 2.0-3.2MM
MDR report key: 2121516
·
Received May 6, 2011
Report
- Report Number
- 1811755-2011-01636
- Event Type
- Malfunction
- Date Received
- May 6, 2011
- Date of Event
- April 12, 2011
- Report Date
- April 12, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PINS ARE STICKING IN THE DEVICE. IT IS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT, BUT THERE WERE NO ALLEGATIONS OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT. NO FURTHER INFORMATION IS AVAILABLE FROM THE ACCOUNT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADJ PIN COLLET 2.0-3.2MM | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |