FDA Adverse Event Malfunction Summary report: N

ADJ PIN COLLET 2.0-3.2MM

MDR report key: 2121516 · Received May 6, 2011

Report

Report Number
1811755-2011-01636
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
April 12, 2011
Report Date
April 12, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PINS ARE STICKING IN THE DEVICE. IT IS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT, BUT THERE WERE NO ALLEGATIONS OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT. NO FURTHER INFORMATION IS AVAILABLE FROM THE ACCOUNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADJ PIN COLLET 2.0-3.2MM KIJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK