FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2121514 · Received June 13, 2011

Report

Report Number
2531779-2011-04083
Event Type
Malfunction
Date Received
June 13, 2011
Date of Event
March 7, 2011
Report Date
May 16, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). CORRECTION/REMOVAL REPORTING NUMBER: 2531779-03/24/2010-003-R. DURING EVALUATION, THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.

Description of Event or Problem · 1

PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 42