FDA Adverse Event Malfunction Summary report: N

MAXIMO VR

MDR report key: 2121481 · Received June 13, 2011

Report

Report Number
6000144-2011-02095
Event Type
Malfunction
Date Received
June 13, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S37
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. WRITE TO LOCKED RAM POR (POWER ON RESET), ADDRESS=10CD, DATA=59 ON (B)(6) 2010 02:12:51. 1 - PATIENT ALERT FOR POR ON (B)(6) 2010 02:12:51.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXPERIENCED AN ELECTRICAL RESET. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAXIMO VR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. 7232CX ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 YR Other 6949 IMPLANTABLE TACHY LEAD