FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2121424 · Received May 6, 2011

Report

Report Number
3002158293-2011-00512
Event Type
Malfunction
Date Received
May 6, 2011
Date of Event
February 18, 2011
Report Date
May 4, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY PACK (B)(4) IS CURRENTLY UNDERWAY. THE REPORTED PROBLEM (BATTERY NOT HOLDING A CHARGE) HAS BEEN CONFIRMED. A ROOT CAUSE ANALYSIS IS CURRENTLY UNDERWAY. A SUPPLEMENTAL REPORT WILL BE SENT UPON COMPLETION OF EVAL. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK.

Description of Event or Problem · 1

A (B)(6) FEMALE PT, CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HONE OF HER BATTERIES WOULD NOT POWER UP A MONITOR. A SUBSEQUENT DOWNLOAD SHOWED THE BATTERY WAS NOT HOLDING A CHARGE. THE PT WAS ISSUED A REPLACEMENT BATTERY PACK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE COARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 72 YR