HEARTSTART MRX
Report
- Report Number
- 1218950-2011-01285
- Event Type
- Malfunction
- Date Received
- May 4, 2011
- Report Date
- April 8, 2011
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
(B)(4). THE CUSTOMER REPORTED THAT THE UNIT FAILED THE OPCHECK WITH A "FAIL/DX." THE CUSTOMER ALSO REPORTED THAT THE UNIT IS NOT RECOGNIZING A NON-PHILIPS BATTERY IN THE B BATTERY COMPARTMENT. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. THE UNIT WAS EVALUATED AT PHILIPS. THE REPORTED ISSUE WITH THE NON-PHILIPS BATTERY COULD NOT BE CONFIRMED AS THE BATTERY WAS NOT SENT IN WITH THE UNIT. UPON EVALUATION, THIS UNIT HAD A CRITICAL CHARGE SHOCK FAILURE. REPLACEMENT OF THE THERAPY PCA RESOLVED THE FAILURE. THE UNIT PASSED ALL POST-SERVICING TESTING AND WAS SENT BACK TO THE CUSTOMER SITE.
THE CUSTOMER REPORTED THAT THE UNIT FAILED THE OPCHECK WITH A "FAIL/DX." THE CUSTOMER ALSO REPORTED THAT THE UNIT IS NOT RECOGNIZING A NON-PHILIPS BATTERY IN THE B BATTERY COMPARTMENT. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTART MRX | MKJ | PHILIPS HEALTHCARE | M3535A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |