FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 3ML LL W/NDL 25X5/8 RB

MDR report key: 21211771 · Received January 22, 2025

Report

Report Number
1213809-2025-00020
Event Type
Malfunction
Date Received
January 22, 2025
Date of Event
January 2, 2025
Report Date
February 7, 2025
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903095705
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

ONE SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. A VISUAL INSPECTION WAS PERFORMED WITH 30X MAGNIFICATION NO DEFECTS OR IMPERFECTIONS WERE OBSERVED. THE SAMPLE WAS ASSEMBLED TO A SYRINGE WITH SALINE SOLUTION. THE SOLUTION WAS EXPELLED WITH NO ISSUED OBSERVED. BASED ON THE INVESTIGATION AND WITH THE SAMPLE ANALYSIS THE SYMPTOM REPORTED COULD NOT BE CONFIRMED. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1161233. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

MATERIAL # 309570, BATCH # 1161233. IT WAS REPORTED THAT THE BD SYRINGE 3ML LL W/NDL 25X5/8 RB HAD A SYRINGE NEEDLE CONNECTIVITY ISSUE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. NEEDLE COMES OFF OF SYRINGE ITEM(S): 309570, LOT(S): 1161233. DATE INCIDENT OCCURRED: ON GOING WAS THE PATIENT IMPACTED? YES, NEEDLE COMES OFF OF SYRINGE IF YES, DESCRIBE: 3 BOTTLES OF B12 WASTED IS A SAMPLE AVAILABLE?: YES. CUSTOMER REQUEST?: NEEDING REIMBURSEMENT FOR SYRINGES AND B12. CUSTOMER RESPONSE ON (B)(6) 2025. THAT¿S WHAT YOU CALL REALLY BAD MULTI-TASKING. SORRY. YES, I WAS WAITING FOR YOU TO CONTACT ME ABOUT THE BAD SYRINGES AND THE LOSS OF THREE BOTTLE SO MY B12 SERUM, WHICH IS COSTLY. THE BAD SYRINGES OCCURRED ON THESE DATES: (B)(6) 24 AND (B)(6) 2025. I DID NOT GET A SHIPPING LABEL AND I¿M NOT SURE WHAT YOU WANT ME TO RETURN. I DISCARDED TWO OF THE THREE BAD SYRINGES. I HAVE THE OPEN AND NOW USELESS B12 BOTTLES AND TWO OF THE PHARMACY BOTTLES.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
520435 BD SYRINGE 3ML LL W/NDL 25X5/8 RB PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 1161233 30382903095705

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown