FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 3ML LL W/NDL 25X5/8 RB

MDR report key: 21211763 · Received January 22, 2025

Report

Report Number
1213809-2025-00021
Event Type
Malfunction
Date Received
January 22, 2025
Date of Event
November 18, 2024
Report Date
February 7, 2025
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903095705
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ONE SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. A VISUAL INSPECTION WAS PERFORMED WITH 30X MAGNIFICATION NO DEFECTS OR IMPERFECTIONS WERE OBSERVED. THE SAMPLE WAS ASSEMBLED TO A SYRINGE WITH SALINE SOLUTION. THE SOLUTION WAS EXPELLED WITH NO ISSUED OBSERVED. BASED ON THE INVESTIGATION AND WITH THE SAMPLE ANALYSIS THE SYMPTOM REPORTED COULD NOT BE CONFIRMED. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1161233. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

MATERIAL # 309570 BATCH # 1161233 IT WAS REPORTED THAT THE BD SYRINGE 3ML LL W/NDL 25X5/8 RB HAD A SYRINGE NEEDLE CONNECTIVITY ISSUE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. NEEDLE COMES OFF OF SYRINGE ITEM(S): 309570 LOT(S): 1161233 DATE INCIDENT OCCURRED: ON GOING WAS THE PATIENT IMPACTED? YES, NEEDLE COMES OFF OF SYRINGE IF YES, DESCRIBE: 3 BOTTLES OF B12 WASTED IS A SAMPLE AVAILABLE?: YES CUSTOMER REQUEST?: NEEDING REIMBURSEMENT FOR SYRINGES AND B12.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
622106 BD SYRINGE 3ML LL W/NDL 25X5/8 RB PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 1161233 30382903095705

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown