HOMECHOICE LOW RECIRCULATION VOLUME APD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-07504
- Event Type
- Malfunction
- Date Received
- June 13, 2011
- Date of Event
- May 19, 2011
- Report Date
- May 19, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K012988
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). A SAMPLE WAS RECEIVED AND EVALUATED. THIS COMPLAINT FOR A DAMAGED LINE WAS CONFIRMED IN THE LAB. SAMPLE EVALUATION REVEALED A CUT APPROXIMATELY 1/2 INCH BEFORE THE TACK WELD ON THE DRAIN LINE. THE ROOT CAUSE IS UNDETERMINED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
BAXTER HOMECARE SERVICES REPRESENTATIVE (HCSR) LEFT A VOICE MESSAGE FOR (B)(6) TO RELAY REPORT RECEIVED FROM THE HOME PATIENT'S MOTHER (HPM) THAT TWO CASSETTES IN WHICH A CRACK/SLASH MARK WAS DETECTED "WHERE THE MACHINE CRIMPS IT" IN TUBE ON UNKNOWN DATE(S). NO PATIENT INVOLVEMENT, INJURY OR ADVERSE EVENT IS REPORTED. THIS IS REPORT 1 OF 2. THE CAREGIVER (CG) WAS CONTACTED ON (B)(6) 2011. THE CG STATED THAT THE MARKS LOOKED LIKE LITTLE SLICES IN THE TUBING CLOSEST TO THE CASSETTE. THE CG WAS NOT SURE IF THE LITTLE SLICES HAD PENETRATED THOUGHT THE TUBING, BUT THEY DID NOT WANT TO USE THOSE CASSETTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE LOW RECIRCULATION VOLUME APD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME | H11A06077 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |